Centurion Range - Ductless Fume Cupboard for Schools

educational-centurion-to-bb88.jpg

The Centurion ductless fume cupboard is suitable for educational institutions, it complies with building bulletin 88. ( Rev to Design Note 29 ) 

Independently Tested by CLEAPPS

Independently Tested to BS EN 14175

The fume cupboard offers teaching staff and students a safe and economical solution to practical science.

The fume cupboard is a stand alone solution complete with its own integral fan and filter system.

The fume cupboard is mounted on a mobile trolley and can be easily moved from room to room for demonstration purposes.

Fume Cupboard Specification

Framework fabricated from powder coated mild steel.

Solid Grade Laminate ( Trespa Athalon ) work base

Hinged 6mm polycarbonate sash front visor.

6mm toughened glass side and rear vision panels.

55W low energy light fitting and light switch

Hour Counter

RCD Protected

The fume cupboard is mounted on a powder coated mild steel mobile storage cupboard, which houses the fan and filter unit.

Easy access for filter replacement.

 

Services Kit ( Optional )

Cold water valve and outlet complete with 2.0mtr flexible hose complete with quick release connection.

Vulcathene drip cup, bottle trap and 2.0mtr of flexible vulcathene drain pipe 

Gas outlet and tap complete with 2.0mtr flexible hose and quick release connection

Docking Station

Robust epoxy coated mild steel lockable enclosure housing all service connections for our Centurion ductless fume cupboard.

The station can be fitted surface mounted or flush into a carcass e.g. Teachers bench or wall.

Note: Special attention must be made when positioning the docking station. The drainage must be lower than the exit drainage from the fume cupboard, in order for waste to drain away efficiently.

Gas and water quick release valves are male and female so service hoses cannot be incorrectly connected.

 1 x electrical socket for power supply to fume cupboard

1 x restraining point for restraining cable from fume cupboard to ensure accidental pulling on service connections.

1 x 38mm vulcathene drain connection with cover when not in use.

1 x quick release female gas valve - 1/4” BSP threaded connection

1 x quick release male cold water valve - 1/4” BSP threaded connection

1 x lockable door

All services clearly labelled

Gas Outlet.

13Amp Double Electrical Socket.

Service control valves are located at the front of the services fascia.

The gas and water services come complete with flexible hoses. ( 2 metres in length ) with quick release connections. 

Dimensions

Centurion10 - 1000mm(w) x 725(d) x 1850(h) fixed or mobile

Centurion12 - 1200mm(w) x 725(d) x 1850(h) fixed or mobile

Centurion15 - 1500mm(w) x 725(d) x 1850(h) fixed or mobile

Performance

Average Face Velocity > 0.3m/sec

Noise Level <50dB

Illumination: >400 Lux

Power Supply 230V

Centrifugal Single Phase Fan Brushless and Sparkless

Filtration

Carbon Grade

G048 (50%)

G048 CT (25%)

G048 HP (35%)

Type: Triple bed (series)

Carbon Weight 25kg

Including pre-filter

Size: 500 x 400 x 100

Volume of air treated 280m3/hour

 

Contaminants

Wide range of inorganics, organics, acids and

Bases. 

Other filters availble on request 

educational-centurion-to-bb88.jpgcenturion-ductless-fume-cupboard-for-schools_73.jpgcenturion-ductless-fume-cupboard-for-schools_13.jpg